Measurement of Self Inductance of High Quality Factor Coil by Hay's Bridge 
Procedure

Fig 1: Measurement of Self Inductance of High Quality Factor Coil by Hay's Bridge
- Connect all the components and the air cored coil as shown in the Fig. 1.
- Set the voltage at a convenient value (say, V=3V) . Set the resistances R2, and R3 at a convenient value and obtain the balance by varying R4 and C4.
- Decide the ranges for R4 and C4 through which they can varied until milivoltmeter (mV) value would be zero.
- Repeat the procedure with different values of the product R2R3 and decide upon readings that permit maximum accuracy for the measurement.