Measurement of Self Inductance of High Quality Factor Coil by Hay's Bridge            

Procedure


Fig 1: Measurement of Self Inductance of High Quality Factor Coil by Hay's Bridge



  1. Connect all the components and the air cored coil as shown in the Fig. 1.
  2. Set the voltage at a convenient value (say, V=3V) . Set the resistances R2, and R3 at a convenient value and obtain the balance by varying R4 and C4.
  3. Decide the ranges for R4 and C4 through which they can varied until milivoltmeter (mV) value would be zero.
  4. Repeat the procedure with different values of the product R2R3 and decide upon readings that permit maximum accuracy for the measurement.